臺灣標準電機SEW儀器儀表--2105 ER Earth Resistance Testers 3 Wire Analog
    <small id="opmff"><tbody id="opmff"></tbody></small>

        1.     
           
          2105 ER   Earth Resistance Testers
          3 Wire (Analog)  
          2105 ER
          FEATURES
           
           Capable of measuring earth voltage.
           2mA measuring current permits earth resistance tests with tripping earth leakage current breakers in the circuit under test.
           Test leads are supplied as standard accessories for simplified two-wire measuring system.
           Battery operated.
           Battery life indication.
           Designed to meet CAT III 30V.
           Calibration performed with supplied test leads.
          SPECIFICATIONS
           
          Measuring Ranges Earth Resistance : 0-12 Ohm/0-120 Ohm/0-1200 Ohm
          Earth Voltage : 0-30V AC (40-500Hz)
          Accuracy Earth Resistance : +/- 3% of full scale
          Earth Voltage : +/- 2.5% of full scale
          Measuring System Earth resistance by constant current inverter (Square Signal) 820 Hz approx. 2mA
          Power Source 1.5 V SUM-3 (R6P) x 8 or equivalent
          Dimensions 210(L) x 210(W) x 100(D) mm
          Weight Approx. 1370g (battery included)
          Accessories Test leads (AL-36 : red-15m, yellow-10m, green-5m)
          Auxiliary earth spikes
          Shoulder belt
          Instruction manual
          batteries
          特 點
          ● 吊線式高品質(zhì)防震錶頭。
          ● 可量測接地電阻(Ω)和接地電壓(ACV)。
          ● 量測電流僅有2mA,不會影響漏電斷路器的跳脫。
          ● 可作三線式量測或兩線式量測。
          ● 電池電量指示。
          ● 使用電池操作。
          ● 請使用提供的測試線測試
          ● 通過歐洲安全規(guī)範驗證:
          EN 61010-1 CAT III 150V
          EN61326-1
          規(guī) 格 (23°C±5°C <80%R.H.)
          量測檔位 接地電阻 : 0-12Ω / 0-120Ω / 0-1200Ω
          接地電壓 : 0-30V AC(40-500Hz)
          準確度 接地電阻 : ±3%滿刻度
          接地電壓 : ±2.5%滿刻度
          量測系列 820Hz 2mA定電流
          接線正確(量測接地電阻) 紅燈(LED)會亮著
          外形尺寸 210(長)×210(寬)×100(高)mm
          重量 約1370公克(含電池)
          使用電池 1.5V(AA)×8(顆)