IRM-5000P Insulation Resistance Meter
----------部分中文介紹產(chǎn)品已停產(chǎn)或參數(shù)有變更,內(nèi)容僅供參考。

IRM-5000P

Insulation Resistance Meter

Part Number(s)
9049-UC [110V]
9050-UC [220V]
Download Links
IRM-5000P Datasheet (English)
Customer References
IRM-5000P User's Manual (English)

Vanguard Instrument Company’s Model IRM-5000P Megohmmeter is a micro-processor based, high-voltage, insulation-test meter. The Model IRM-5000P is rugged and portable, making it ideal for use in electric-utility substations and for industrial application.

This sophisticated insulation tester uses a dual-microprocessor design. One microprocessor is dedicated to the control of the power supply and measuring circuitry. The second microprocessor is dedicated to the display, printer control, and user interfaces.

  • Dual-processor design
  • Perform PI, SV & DD tests automatically
  • Test duration from 1 to 90 minutes
  • Automatically discharge test voltage
  • Measures capacitance of DUT
  • Store up to 100 test reports
  • IR temperature probe, provide equivalent resistance reading at 40°C/10°C
  • Built-in 2.5" thermal printer
  • Computer Interface

IRM-5000P

Cables

IRM-5000P

Technical Specifications

TYPE Insulation resistance meter
PHYSICAL SPECIFICATIONS 19”W x 7”H x 15”D, (48 cm x 17 cm x 38 cm); Weight: 24lbs (10.9 kg)
INPUT POWER 110 – 120 Vac or 220 – 240 Vac (selectable), 50/60Hz
RESISTANCE READING RANGE
(0°C to +30°C)
100 K-ohms – 1 M-ohm ±20%,
1 M-ohm – 1 T-ohm ±5%,
1 T-ohm – 5 T-ohm ±20% , 5kV
100 K-ohms – 1 M-ohm ±20%,
1 M-ohm – 100 G-ohms ±5%,
100 G-ohms – 500 G-ohms ±5%, 500 V
1 M-ohm – 10 G-ohms ±5%, 50 V
TEST VOLTAGE Selectable from 50Vdc – 5 KVdc, in 2 Vdc steps
OUTPUT VOLTAGE ACCURACY
(0°C to +30°C)
±2%, ±1v of selected voltage with load resistance greater than 100 MΩ
SHORT CIRCUIT CURRENT 2 mA max
CURRENT READING RANGE
(0°C to +30°C)
0.03 nA – 2 mA; Accuracy: ±5% ±0.2 nA
CAPACITANCE READING RANGE
(0°C to +30°C)
0.01 μF – 10.0 μF (Test voltage greater than 100V); Accuracy: ±5% ±0.03 μF
VOLTAGE READING RANGE
(0°C to +30°C)
50 – 1250 Vac (rms) or dc; Accuracy: ±5%, ±1V
CAPACITOR DISCHARGE Less than 2 Sec/μF, automatic at the end of test
HUM REJECTION 1mA per 1kv of test voltage, 2mA rms maximum
BATTERIES Two 12V, 2.0 Ah sealed lead acid batteries. Battery life: typical 6 hrs, continuous testing. The IRM-5000P can be used while charging.
PRINTER Built-in 2.5-inch wide thermal printer can print test results in both tabulated and graphic formats
COMPUTER INTERFACE One RS-232C port
SAFETY
ENVIRONMENT Operating: -10 to 50 C (15 to +122 F); Storage: -30 C to 70 C (-22 to +158 F)
CABLES One 6-foot cable set, one 15-foot cable set, ground cable, power cord, cable bag
OPTIONS Transportation Case
WARRANTY One year on parts and labor
Specifications are valid at nominal voltage and ambient temperature of +25C (+77F).
Last updated on February 6, 2009

Field Pictures

IRM-5000P at Rotex Engineering, South Africa